Oxygen Delivery

Medical Device Calibration and Functional Test
Medical device (Demand Oxygen Delivery System) is calibrated and functionally tested at a rate of 18 parts per hour.

Calibration includes:
• Sensitivity Calibration - monitoring the pulse volume while adjusting the sensitivity potentiometer.
• Manifold Calibration - monitoring the pulse volume while the flow is manually adjusted.

Functional test includes:
• Auto triggering test (ensures that the device does not trigger unintentionally)
• Pulse verification test (checks all pulse settings to meet required volume limits)
• Continuous flow test (user adjustable, customer specified operating flow range, 1.8 – 2.2 lpm typical)
• Low battery test (check sensitivity and pulse volume are within customer specified limits, 13 – 80 ml/pulse typical)

Optional – Apnea test