Pressure Decay
M1045z/M1075z Series Pressure Change Leak Detector
The InterTech M1045z Pressure Change or M1075z Differential Pressure Change Leak Detectors are a low cost, high resolution, pressure change instruments suitable for a wide range of part volumes.


M1045-03z Pressure Change Leak Detector
Pressure change testing, compact construction, programmable pressure regulator, and high (24-bit) resolution instrumentation for small volume parts.
The pressure transducer is connected to the test part. When this circuit is filled with air and isolated from the supply, any leak results in a pressure change relative to atmospheric pressure. The transducer output provides a measurement of the pressure change proportional to the leak rate.
• Programmable Pressure Range: 0 - 150 psig
• Transducer repeatability: 0.05% F.S.D.
• Transducer sensitivity: < 10-5 psi (STD)
• A/D Conversion: 24 bit, 100 conversions/second
• Timer increments: 0.01 sec.
• Low (<1.5 cm3) internal volume

M1045-19z Pressure Change Leak Detector
Pressure change testing, compact construction, programmable pressure regulator, and high (24-bit) resolution instrumentation for wide range of part volumes.
The pressure transducer is connected to the test part. When this circuit is filled with air and isolated from the supply, any leak results in a pressure change relative to atmospheric pressure. The transducer output provides a measurement of the pressure change proportional to the leak rate.
• Programmable Pressure Range: 0 - 150 psig
• Transducer repeatability: 0.05% F.S.D.
• Transducer sensitivity: < 10-5 psi (STD)
• A/D Conversion: 24 bit, 100 conversions/second
• Timer increments: 0.01 sec.
• Low (<1.5 cm3) internal volume

M1075-26z Differential Pressure Decay
The InterTech M-1075-26z Leak Detector is a high pressure construction, high resolution, differential pressure change instrument suitable for a medium to large part volumes.
Test item and reference volume are pressurized and then isolated by closing valve 1. Reference volume is isolated from test item by closing valve 2. Transducer reads pressure differential between reference volume and test item twice over time.
• Pressure Range: 0 - 450 psig (standard)
• Transducer Range: 0 - 10 In-WC (standard)
• Transducer repeatability: 0.05% F.S.D.
• Transducer sensitivity: < 10-5 psi (STD)
• A/D Conversion: 24 bit, 100 conversions/second
• Timer increments: 0.01 sec.
• Low (<1.5 cm3) internal volume


InterTech Development Company 7401 N. Linder Avenue, Skokie, IL 60077, USA T. 847.679.3377
InterTech Europe 19, rue Clement Michut F-69100 Villeurbanne,France (33)0437.434.116
Shanghai Technical Center 1F, Bldg #36, 458 North Fu Te Road Shanghai Waigaoqiao F.T.Z. 200131 P.R. China, (8621)58682809

 
   
19 rue Clément Michut
69100 VILLEURBANNE
Tel : +33 (0)4 37 43 41 16
Fax : +33 (0)4 37 43 41 17
intertechinfo@intertecheurope.com
  InterTech Development Co
7401 N. Linder Avenue
Skokie, 60077 Illinois USA
TEL 847 -679-3377
FAX 847-679-3391
Service : info@intertechdeveloppement.com
Sales : sales@intertechdeveloppement.com
 
HOME | NEWS&EVENTS | CONTACT US